Monochromatic wavelength dispersive x-ray fluorescence (MWD XRF) is an enhanced version of conventional wavelength-dispersive X-ray spectroscopy (WDXRF) elemental analysis. The key difference is that MWD XRF uses a doubly curved crystal X-ray optic between the X-ray source and the sample resulting in monochromatic excitation. Monochromatic wavelength dispersive x-ray fluorescenceG. Havrilla, et al., "Monochromatic Wavelength Dispersive X-Ray Fluorescence for Improved Pu Assay" International Atomic Energy Agency Safeguards Symposium, November 2010 Vienna, Austria This additional optic creates a high-intensity X-ray beam on a small spot size without increasing the power of the X-ray source. An MWD XRF instrument is constructed from a low-power X-ray tube, a point-to-point focusing optic for excitation, a sample cell, a focusing optic that collects the fluorescence from the sample, and an X-ray detector. By using an optic between the X-ray source and the sample, a monochromatic beam free of bremsstrahlung, excites the sample, eliciting the secondary fluorescence X-rays needed for elemental analysis. By restricting the band of wavelengths used for excitation, a much higher signal to background ratio is achieved. This type of excitation allows much lower limits of detection and faster reading times.Z. W. Chen, et al., Journal of ASTM International September 2005 Volume 2 No. 8