Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a mass spectrum, a plot of intensity as a function of the mass-to-charge ratio. These spectra are used to determine the elemental or isotopic signature of a sample, the masses of particles and of molecules, and to elucidate the chemical identity or structure of molecules and other chemical compounds. Mass spectrometry is used in many different fields and is applied to pure samples as well as complex mixtures.

In a typical MS procedure, a sample, which may be solid, liquid, or gaseous, is ionized, for example by bombarding it with a beam of electrons. This may cause some of the sample's molecules to break up into positively charged fragments or simply become positively charged without fragmenting. These ions (fragments) are then separated according to their mass-to-charge ratio, for example by accelerating them and subjecting them to an electric or magnetic field: ions of the same mass-to-charge ratio will undergo the same amount of deflection. The ions are detected by a mechanism capable of detecting charged particles, such as an electron multiplier. Results are displayed as spectra of the signal intensity of detected ions as a function of the mass-to-charge ratio. The atoms or molecules in the sample can be identified by correlating known masses (e.g. an entire molecule) to the identified masses or through a characteristic fragmentation pattern.

History

In 1886, Eugen Goldstein observed rays in gas discharges under low pressure that traveled away from the anode and through channels in a perforated cathode, opposite to the direction of negatively charged cathode rays (which travel from cathode to anode). Goldstein called these positively charged anode rays "Kanalstrahlen"; the standard translation of this term into English is "canal rays". Wilhelm Wien found that strong electric or magnetic fields deflected the canal rays and, in 1899, constructed a device with perpendicular electric and magnetic fields that separated the positive rays according to their charge-to-mass ratio (Q/m). Wien found that the charge-to-mass ratio depended on the nature of the gas in the discharge tube. English scientist J. J. Thomson later improved on the work of Wien by reducing the pressure to create the mass spectrograph.

The word spectrograph had become part of the international scientific vocabulary by 1884. Early spectrometry devices that measured the mass-to-charge ratio of ions were called mass spectrographs which consisted of instruments that recorded a spectrum of mass values on a photographic plate. A mass spectroscope is similar to a mass spectrograph except that the beam of ions is directed onto a phosphor screen. A mass spectroscope configuration was used in early instruments when it was desired that the effects of adjustments be quickly observed. Once the instrument was properly adjusted, a photographic plate was inserted and exposed. The term mass spectroscope continued to be used even though the direct illumination of a phosphor screen was replaced by indirect measurements with an oscilloscope. The use of the term mass spectroscopy is now discouraged due to the possibility of confusion with light spectroscopy. Mass spectrometry is often abbreviated as mass-spec or simply as MS.

Modern techniques of mass spectrometry were devised by Arthur Jeffrey Dempster and F.W. Aston in 1918 and 1919 respectively.

Sector mass spectrometers known as calutrons were developed by Ernest O. Lawrence and used for separating the isotopes of uranium during the Manhattan Project. Calutron mass spectrometers were used for uranium enrichment at the Oak Ridge, Tennessee Y-12 plant established during World War II.

In 1989, half of the Nobel Prize in Physics was awarded to Hans Dehmelt and Wolfgang Paul for the development of the ion trap technique in the 1950s and 1960s.

In 2002, the Nobel Prize in Chemistry was awarded to John Bennett Fenn for the development of electrospray ionization (ESI) and Koichi Tanaka for the development of soft laser desorption and their application to the ionization of biological macromolecules, especially proteins.

Components

A mass spectrometer consists of 5 components: sample inlet, ion source (or ionizer), mass analyzer, detector, data system. Each component can be varied somewhat independently from the other components, and the full mass spectrometer itself can be included as a component in an analysis pipeline, allowing great flexibility.

The sample inlet prepares the sample to be analyzed in a suitable state.

The ion source takes samples from the sample inlet, converts a portion of the sample into a stream of ions, and outputs it to the mass analyzer. There is a wide variety of ionization techniques, depending on the phase (solid, liquid, gas) of the sample and the efficiency of various ionization mechanisms for the unknown species. An extraction system removes ions from the sample, which are then targeted into the mass analyzer.

The mass analyzer takes an input stream of ions and separates ("analyzes") its constituent ions according to their m/z, the mass-to-charge ratio.

The detector measures the value of an indicator quantity and thus provides data for calculating the abundances of each ion present. Some detectors also give spatial information, e.g., a multichannel plate.

Illustrative example The following describes the operation of a simple mass spectrometer that uses sector type. Other types are treated below.

Consider a sample of sodium chloride (table salt). In the ion source, the sample is vaporized (turned into gas) and ionized (transformed into electrically charged particles) into sodium (Na+) and chloride (Cl−) ions.

Sodium atoms and ions are monoisotopic, with a mass of about 23 daltons (symbol: Da or older symbol: u). Chloride atoms and ions come in two stable isotopes with masses of approximately 35 u (at a natural abundance of about 75 percent) and approximately 37 u (at a natural abundance of about 25 percent).

The analyzer part of the spectrometer contains electric and magnetic fields, which exert forces on ions traveling through these fields. The speed of a charged particle may be increased or decreased while passing through the electric field, and its direction may be altered by the magnetic field.

The magnitude of the deflection of the moving ion's trajectory depends on its mass-to-charge ratio. Lighter ions are deflected by the magnetic force to a greater degree than heavier ions (based on Newton's second law of motion, F = ma).

The streams of magnetically sorted ions pass from the analyzer to the detector, which records the relative abundance of each ion type. This information is used to determine the chemical element composition of the original sample (i.e. that both sodium and chlorine are present in the sample) and the isotopic composition of its constituents (the ratio of 35Cl to 37Cl).

Sample inlet The sample inlet provides a stream of neutral molecules, so that the ion source can ionize them. There are several classes.

The classical design produces the stream of molecules in the form of a low-pressure gas. It is low-pressure, so that molecules within the stream do not frequently collide with each other. Ions must travel without frequent collisions, because the instrument separates ions by their motion in electric and magnetic fields. Collisions disturb that motion.

Despite being low-pressure, the sample stream is still at a much higher pressure than the rest of the system, which is held in a greater vacuum. Consequently, the sample stream can only enter the ion source chamber via a small pinhole, called a molecular leak.

If the sample itself is already a gas, then the sample inlet can simply be a gas inlet with pressure valve. If the sample is a volatile liquid or solid, then it can simply be placed in a chamber, vacuum-pumped to a low pressure. If the sample is not sufficiently volatile, the sample inlet can be fitted within an oven. Be careful not to decompose the sample.

If the sample is too nonvolatile to be turned into a low-pressure gas, then one can use a direct probe. The direct probe is a probe with a tip of thin wire loop, pin, or a capillary tube. The sample is placed in the tip. The probe is inserted through a vacuum lock into the ionization chamber, so that the tip is close to the ion source. The probe can be heated to increase vapor pressure further. Such a system is effective for samples with vapor pressures lower than mmHg at room temperature.

The sample can be prepared by an earlier analyzer. For example, it can simply be the gaseous output of a gas chromatography machine as in GC-MS, or the liquid output of a high-performance liquid chromatography machine, as in HPLC-MS.

The recently developed "atmospheric pressure ionization" techniques allow us to entirely dispense with the sample inlet.

Ion source

The ion source is the part of the mass spectrometer that transforms the sample from the sample inlet into a stream of ions. There are many techniques with various dimensions for tradeoffs between different choices of ion sources, in terms of monetary cost, accuracy, precision, compatibility with other stages of spectroscopy, ease of use, etc.

For example, electron ionization (EI) gives a high degree of fragmentation, yielding highly detailed mass spectra which when skilfully analysed can provide important information for structural elucidation/characterisation and facilitate identification of unknown compounds by comparison to mass spectral libraries obtained under identical operating conditions. However, EI is not suitable for coupling to HPLC, i.e. LC-MS, since at atmospheric pressure, the filaments used to generate electrons burn out rapidly. Thus EI is coupled predominantly with GC, i.e. GC-MS, where the entire system is under high vacuum.

Two techniques often used with liquid and solid biological samples include electrospray ionization (invented by John Fenn) and matrix-assisted laser desorption/ionization (MALDI, initially developed as a similar technique Soft Laser Desorption) by K. Tanaka for which a Nobel Prize was awarded and as MALDI by M. Karas and F. Hillenkamp).

Hard ionization and soft ionization

Hard ionization techniques are those that impart high quantities of residual internal energy in the subject molecule, even after ionization. Part of the energy would be dissipated by rupturing chemical bonds within the ion, producing many fragmentary radical cations. The resultant ions tend to have m/z lower than the molecular ion (other than in the case of proton transfer and not including isotope peaks). The most common example of hard ionization is electron ionization (EI).

The radical ions produced after the first ionization are called second-generation product ions. The products are then directed towards the mass analyzer by a repeller electrode. The ionization process often follows predictable cleavage reactions that give rise to fragment ions which, following detection and signal processing, convey structural information about the analyte.

Conversely, soft ionization techniques are those that impart little residual energy onto the subject molecule and as such result in little fragmentation. Examples include fast atom bombardment (FAB), chemical ionization (CI), atmospheric-pressure chemical ionization (APCI), atmospheric-pressure photoionization (APPI), electrospray ionization (ESI), desorption electrospray ionization (DESI), and matrix-assisted laser desorption/ionization (MALDI).

Electron ionization